4.2 Zeroing the Gauge
The gauge is factory calibrated while "standing upright"
( "Calibration Test Report").
We recommend performing a zero adjustment, when the
gauge is operated for the first time.
Due to long time operation or contamination, a zero drift could
occur and zero adjustment may become necessary.
For adjusting the zero, operate the gauge under the same con-
stant ambient conditions and in the same mounting orientation
as normally.
The output signal (measuring signal) is depending on the moun-
ting orientation. The signal difference between the vertical and
horizontal mounting orientation is:
F.S. ΔU / 90°
1000 Torr/mbar 2 mV
100 Torr/mbar 10 mV
10 Torr/mbar 50 mV
1 Torr/mbar 300 mV
If the gauge is operated via a controller, the zero of the
whole measuring system has to be adjusted on the
controller: first, adjust the zero of the gauge and then,
the zero of the controller.