tqna49e1 (2010-05) 21
4.2.1 <ZERO> Adjustment
n Evacuate the gauge to a pressure according to the table
below:
F.S.
Recommended final pressure for
zero adjustment
1000 Torr/mbar
100 Torr/mbar
10 Torr/mbar
1 Torr/mbar
<5
×10
-2
Torr
<5
×10
-3
Torr
<5
×10
-4
Torr
<5
×10
-5
Torr
<6.65
×10
0
Pa
<6.65
×10
-1
Pa
<6.65
×10
-2
Pa
<6.65
×10
-3
Pa
<5
×10
-2
mbar
<5
×10
-3
mbar
<5
×10
-4
mbar
<5
×10
-5
mbar
If the final pressure in the gauge is too high for zero ad-
justment (>25% of the F.S.), the zero cannot be reached
and the <RUN> LED flashes. If this is the case, activate
the factory setting and adjust the zero again ( 27).
o Operate the gauge for at least ¼ hour (until the signal is
stable).
p Briefly press the <ZERO> button with a pin (max.
ø1.1 mm). The zero adjustment runs automatically. The
<STATUS> LED flashes until the adjustment (duration
8 s) is completed.
Press the button briefly
max. ø1.1 mm